Loading article...
Loading article...
Stream/Area:
-
Subject:
-
Paper ID:
10.55041/IJSREM21948
Volume:
7 |
5 |
May 2023
Pages:
Publication Frequency:
Monthly (12 issues per year)
License:International Journal of Scientific Research in Engineering and Management follows the Creative Commons Attribution 4.0 (CC BY 4.0) License, allowing free use and distribution with proper attribution.
A simple and specific method for determination of bioavailability of Mefenamic acid/ Meloxicam/ Aceclofenac in Human Plasma by liquid chromatography-tandem mass spectrometry operated in positive ionization mode was developed and validated. The LC-MS/MS methods were highly sensitive and suitable for the detection of drug in plasma even in low concentrations. Solid phase extraction methodology was adopted in plasma sample preparation, which provides consistent extraction recovery with minimal endogenous interference and matrix effect. The method involves a rapid solid phase extraction from plasma, simple isocratic chromatographic conditions and mass spectrometric detection in the positive ionization mode using multiple reactions monitoring that enables detection down to low nanogram levels with a short run time. Different parameters such as linearity, range, precision, accuracy, ruggedness and robustness, limit of detection (LOD) and limit of quantification (LOQ) were used for full validation of the method. The results were found to be acceptable as per the guidelines of International Conference on Harmonization (ICH). The method found to be novel, rapid, linear, precise, accurate, robust and rugged and can be successfully applied for the routine analysis of Mefenamic acid / Meloxicam/ Aceclofenac with more sensitivity and covers wider range of quantitation. The method also found to be useful and economical.
KEYWORDS
Publication Date
5/25/2023
Peer Review Type
Double-Blind Peer Review
Review Rounds: 2
Documents
Article Views
0
Total views since publication
PDF Downloads
1
Total downloads since publication
Geographic Reach
5
Digital Preservation: This article is permanently archived on Zenodo and complies with all data preservation standards across major indexing platforms.
Plagiarism Screening
This manuscript has been checked for plagiarism using TurnitinPlagiarismChecker software.
Conflict of Interest
The authors declare no conflict of interest related to this research work.
Funding Information
No funding information available for this research.